英语翻译Completion of the calibration by means of the TRIM functionAfter an operating time of several days,the moisture value should be determined once again byway of a laboratory analysis.If the laboratory values are consistent with the measured

来源:学生作业帮助网 编辑:作业帮 时间:2024/12/01 13:51:48
英语翻译Completion of the calibration by means of the TRIM functionAfter an operating time of several days,the moisture value should be determined once again byway of a laboratory analysis.If the laboratory values are consistent with the measured
xY]O+Te%[W",j#%hHW1b<D 6!@`L` '9q9xd&j+Y}3n~q'M}(WS`zUQZȥ6* {KST|b!UȚ},YT1CBqHTP}%V9Q֯ !-FLP4awX@ +:uc[Mۖ5 =Ha-5(% ˅ȁ>N6|(.i'n@I=T$o71Cúq5Ɇ LoF 21=ׇ Ai b԰0`6 |;)w_|'X{jղ4@._Ԉb|Ӗ{CMJJo%gksHrWzL#շݛ -#3 a8ŗV7H:̔H*F]ɨ ʐiX#p_iJ;=w=u7ncoW_{dd4sWOOm!)-.e oiMdOOD+#E)@aVuup6b=}e_wL;!F7E膁ViU=nI"'{}Wz[]1~%rwoz 0pZ$Ku%*6Kh9atYTCV:2-bBk)@'nS"jNQi#K @cw{żw>\pV0wt\ 㭣.OZ} Ä䊎/KV#)5Y'hcO*HxvMZL5Ykǔ'[kvcׯ5KÊ>Аkz _Ej]%Y~ޓ]j rYIg[h=WҘYvY{Wq6xv*OݵB:) <7a? }^Y3a9f?-"uՃ`?,гelgT=En!K0gՊ9Gi<}RmVΆ"g sj?> u.j/gpW"@I.Tb=![3] 8g@$Uosv0R{ T(i#CP%T1D*~gN1$6!Xzj>;jF>FNy_5{d?Lϐ??VmNyzjtOn]7$ڋo>3M'{ ܀^&.>x rf % >Dˌu"ܤl\}kҎu}N} ̽ ᙼD!Kq=y!X <87QF`[#M9[y^/Q~Xӱ@wQvզg9S^fοuo%J)]K@^K|CK1&A.AɲTB/,]'O߽ހ&oLh_bXtH_T~> wc#=fHԋa?C<, S,~=j@ nxf'Tml١ HK-Ԫh x"b,tIdgB^zmԂ!t T4޸Wg9)wdUT"L IA(Ra%YEύ)L 6 !Ncz,vy$(@>TY#UاWQο:+k

英语翻译Completion of the calibration by means of the TRIM functionAfter an operating time of several days,the moisture value should be determined once again byway of a laboratory analysis.If the laboratory values are consistent with the measured
英语翻译
Completion of the calibration by means of the TRIM function
After an operating time of several days,the moisture value should be determined once again by
way of a laboratory analysis.If the laboratory values are consistent with the measured value of the
moisture gauge,you can abstain from the work step which is described in the following.If the
difference between the values is very high,the wort step Ñoptimization of the calibration by the
procedure of collecting samplesì should be repeated.In case of a small difference between the
measured values,the calibration with the TRIM function can be opted once again.
ï Determine the new actual value by means of the diagram shown below.
ï Press the MENU key.
ï Press the CALIBRATE key.
ï Press the TRIM key.
ï Press the CHANNEL key several times until the required measurement is displayed beside
the CHANNEL key.
ï Press the key NEW ACTUAL VALUE.The DATA ENTRY PAGE is displayed.Enter now the
corrected value.
ï Press the key CURRENT ACTUAL VALUE.The DATA ENTRY PAGE is displayed.Enter now
the current value.The current value is shown below on the display.
ï Press ENTER.
ï This way,the new TRIM value is loaded into the gauge.This new TRIM value is saved also in
the recipe which has currently been selected in the operator interface.After the successful
transfer,the RUN Mask with the newly calculated measured value is displayed again.
ï Repeat this procedure for the other measurements in the product.
" With the TRIM function,you determine the required parallel displacement
(the offset) of the actual graph of the measured values (graph of the current
calibration) towards the graph of the optimum calibration.

英语翻译Completion of the calibration by means of the TRIM functionAfter an operating time of several days,the moisture value should be determined once again byway of a laboratory analysis.If the laboratory values are consistent with the measured
我自己来手动翻译下,机器翻的南面会有一些语法错误,谢谢采纳,有点长哦,呵呵!
通过整理完成校对.通过几天的实行,以及实验分析,决定湿度值.如果实验室的水湿度值和湿度表测量出来的是一致的,你可以不必做接下来的描述的步骤,如果是不同的,而且很高,那么校准以及搜集样本的步骤就要重复进行.万一测量的值只有一点点差别,那么你就再重新整理校对.我通过如下图所示的方法确定实际值.我按下菜单键,接着按下校正键,再按下整理键,按着CHANNEL键直到需要的测量值显示在CHANNEL键旁边.按下实际值的键,数据输入页显示了,现在进入了修正价值.按下CURRENT ACTUAL VALUE键,数据页显示了.新的价值在下面显示.按下进入键.这样,新的整理值会进入表内,也会被保存在操作员接口处.通过成功的传送, RUN Mask 新的积累的数据会重新显示.重复其他测量的步骤.通过整理的和精准的校正功能,你会确定出需要的平行的排水位如实际的测量值.
我一个一个字打得很辛苦的,给个满意答案哈. 上面的回答一看就是在网上翻译的,不过那样一句一句的翻译也很辛苦的.

英语原文:Completion of the calibration by means of the TRIM function
After an operating time of several days, the moisture value should be determined once again by
way of a laboratory analysis. If...

全部展开

英语原文:Completion of the calibration by means of the TRIM function
After an operating time of several days, the moisture value should be determined once again by
way of a laboratory analysis. If the laboratory values are consistent with the measured value of the
moisture gauge, you can abstain from the work step which is described in the following. If the
difference between the values is very high, the wort step Ñoptimization of the calibration by the
procedure of collecting samplesì should be repeated. In case of a small difference between the
measured values, the calibration with the TRIM function can be opted once again.
ï Determine the new actual value by means of the diagram shown below.
ï Press the MENU key.
ï Press the CALIBRATE key.
ï Press the TRIM key.
ï Press the CHANNEL key several times until the required measurement is displayed beside
the CHANNEL key.
ï Press the key NEW ACTUAL VALUE. The DATA ENTRY PAGE is displayed. Enter now the
corrected value.
ï Press the key CURRENT ACTUAL VALUE. The DATA ENTRY PAGE is displayed. Enter now
the current value. The current value is shown below on the display.
ï Press ENTER.
ï This way, the new TRIM value is loaded into the gauge. This new TRIM value is saved also in
the recipe which has currently been selected in the operator interface. After the successful
transfer, the RUN Mask with the newly calculated measured value is displayed again.
ï Repeat this procedure for the other measurements in the product.
" With the TRIM function, you determine the required parallel displacement
(the offset) of the actual graph of the measured values (graph of the current
calibration) towards the graph of the optimum calibration.
中文翻译:完成校准通过修剪功能
一个操作时间后几天,水分值应该是确定的,再一次
一个实验室分析。如果实验值符合的测量值
水分仪,你可以自制的工作步骤描述如下。如果
值之间的差别是非常高的,Noptimization麦汁一步的工作的
收集samplesi程序应该重复。如果一个小差异的
测量值、校准与修剪函数可以选择一次。
我决定采用新的实际价值的图显示如下。
我按下菜单的按键。
我按了标定的钥匙。
我按了修剪的钥匙。
我按了几次通道关键,直到所需要的测量显示身边
频道的钥匙。
我按了重点新实际价值。数据进入页面显示出来。现在进入了
修正的价值。
我按了关键当前实际价值。数据进入页面显示出来。现在进入
当前的价值。当前值显示如下图所示。
我按回车。
我这个途径,新修的价值是加载到表。这个新修的价值是救了还在
目前的配方选择在操作界面。在成功的
转让、运行面膜测量值的新计算显示了。
我重复这个过程测量中的其他产品。
“与修剪功能,你决定所需的平行位移
(位移)的实际测量值的图表(图水流的影响
向图校准)的最佳校准

收起

英语翻译Incentives Dept.(BOI) BOI应该是投资委员会Estimated computation of D/TList of existing machinery and equipment of Exemption and Endorsement Letter W/ same function or use.IMPORTATION OF BOI REGISTERED COMPANIESUpon completion of th certificate of completion stage of completion是什么意思 certificate of completion是什么意思 certificate of completion stage of completion是什么意思 英语翻译The rights and obligations of the parties do not merge on completion of any transaction contemplated by this Agreement and all representations and warranties survive completion. completion completion 英语翻译37.0 Machine bases..After completion of the blinding layer under the machine base,construct a 115mm brick or 100mm concrete block wall around the edge of the base as permanent formwork,and carry this up to the level of the underside of th 翻译a sense of successful completion 英语翻译1 Princeple Cashflow2 No of Days to Completion Date3 No of years to Completion Date4 One Year Interest Rate5 Interest for each installment (annualy compond) in integral years6 Interest for each installment prorata in rest time7 Agreegate 英语翻译(d) The terms,provisions,representations and warranties contained in this Agreement that by their sense and context are intended to survive the completion of performance hereunder by either or both parties shall so survive the completion 英语翻译the provision of in-house engineering capacity,enabling completion of clients’ priority prototype conversion projects in a fraction of the normal time. 英语翻译Construction of a new PVC stabilisers plant in Cilegon,Java is now a year behind the original schedule of a March 2010 completion. 英语翻译While you have not met the English language proficiency requirements of the program to date,you will meet this condition upon completion of your Diploma. 英语翻译不好意思,点错了没把问题写清楚哦,呵呵:Series A Original Issue DateMaterial Adverse EffectMemorandum of AssociationAccelerated Redemption Triggering EventAffiliateRestructuring Completion Dateregistered officeRelated 英语翻译Completion of the calibration by means of the TRIM functionAfter an operating time of several days,the moisture value should be determined once again byway of a laboratory analysis.If the laboratory values are consistent with the measured